  specifications features data sheet mar. 15, 1999 date fha, fka ... 3 fcx ?ax series EDSX5-71 the fcx ax absolute pressure transmitter accurately measures absolute pressure and transmits a proportional 4 to 20ma signal. the transmitter utilizes a unique micromachined capaci- tance silicon sensor with state-of-the-art microprocessor tech-nology to provide exceptional performance and func- tionality. functional specifications type: model fha: 4 to 20ma model fka: 4 to 20ma with digital signal service: liquid, gas, or vapour span, range, and overrange limit: remark: to minimize environmental influence, span should be greater than 1/40 of the max. span in most applications. the maximum span of each sensor can be converted to different units using factors as below. 1mpa abs=10 3 kpa abs=10bar abs=10.19716kgf/cm 2 abs =145.0377psi abs 1kpa abs=10mbar abs=101.9716mmh 2 o abs= 4.01463inh 2 o abs=7.50062mmhg abs output signal: model fha: 4 to 20ma dc 2-wire model fka: 4 to 20ma dc with digital signal super- imposed on the 4 to 20ma signal. power supply: transmitter operates on 10.5v to 45v dc at transmitter terminals. 10.5v to 32v dc for the units with op- tional arrester. f a 01 f a 02 f a 03 f a 04 overrange limit [mpa] {bar} type 1. high accuracy 0.2% accuracy for all calibrated spans is a standard fea- ture for all ap models covering 1.6kpa{0.016bar} range to 3000kpa{30bar} high pressure range. 0.1% accuracy is available as option. fuji's micro-capacitance silicon sensor assures this accuracy for all suppressed calibra- tion ranges without additional adjustment. 2. minimum environmental influence the "advanced floating cell" design which protects the pressure sensor against changes in temperature, and overpressure substantially reduces total measurement error in actual field applications. 3. replaceable communication module fuji micro-electronics manufacturing technology offers replaceable communication module that makes fcx ax transmitter very unique in design. in case of change in communication protocol, all that needs to be done is just to replace the module and the transmitter gets upgraded to the new version. 4. fuji/hart bilingual communication module the communication module is "bilingual" to speak both fuji proprietary protocol and hart. any hart compatible devices can communicate with fcx ax series transmitters. 5. application flexibility various options that render the fcx ax suitable for almost any process applications include: analog indicator at either the electronics side or ter- minal side full range of hazardous area approvals built-in rfi filter and lightning arrester ? -digits lcd meter stainless steel electronics housing wide selection of materials 6. burnout current flexibility (under scale: 3.2 to 3.8ma, over scale: 20.8 to 21.6ma) burnout signal level is adjustable using model fxw hand held communicator (hhc) to comply with namur ne43. (available for amplifier unit from version 24 and fxw(hhc) version 5.3.) 7. dry calibration without reference pressure thanks to the best combination of unique construction of mechanical parts (sensor unit) and high performance electronics circuit (electronics unit), reliability of dry calibration without reference pressure is at equal level as wet calibration. 0.5 {5} 0.5 {5} 1.5 {15} 9 {90} min. range limit [kpa abs] {bar abs} span limit [kpa abs] {bar abs} 1.6 {0.016} 13 {0.13} 50 {0.5} 300 {3} 1.6 {0.016} 1.6 {0.016} 5 {0.05} 30 {0.3} 16 {0.16} 130 {1.3} 500 {5} 3000 {30} max. fha fka fha/fka 0 to +16 {0 to +0.16} 0 to +130 {0 to +1.3} 0 to +500 {0 to +5} 0 to +3000 {0 to +30} absolute pressure transmitter 1 2
fha, fka ... 3 2 load limitations: see figure below ?utput overscale? approx. 21.6ma (adjustable within the range 20.8ma to 21.6ma from hhc) ?utput underscale? approx. 3.8ma (adjustable within the range 3.2ma to 3.8ma from hhc) 0 3.8 3.2 4 20 20.8 21.6 over scale burnout probable over range probable under range normal operating range under scale burnout loop-check output: model fha: transmitter can output constant signal of 4ma, 12ma, or 20ma if mode switch is set to the loop check mode. model fka: transmitter can be configured to provide constant signal 3.8ma through 21.6ma by hhc. temperature limit: ambient: ?0 to +85 c (?0 to +80 c for lcd indicator) (?0 to +60 c for arrester option) for explosionproof units (flameproof or intrinsic safety), ambient temperature must be within the limits specified by each standard. process: 40 to +85 c for silicone fill sensor storage: 40 to +90 c humidity limit: o to 100% rh communication: (model fka only) with hhc (model fxw, consult data sheet no. eds8-47), following informa- tion can be remotely displayed or recon- figured. note: for communication with hhc (model: fxw), min. of 250 ? is required. hazardous locations: (approval pending) type n nonincendive authorities flameproof intrinsic safety items display set tag no. model no. serial no. engineering unit range limit measuring range damping output mode burnout direction adjustment output adjust data self diagnoses printer external switch lock transmitter display(*) vv vv v vv v vv vv vv vv vv ? v v vv vv zero/span adjustment: model fha: zero is adjustable from the external adjustment screw. the adjustment screw can also function to adjust span when mode switch (located on the electronics unit) is in the span mode. inhibit mode to disable the adjustment screw is also available. model fka: zero and span are adjustable from the hhc. zero is also adjustable externally from the adjustment screw. damping: adjustable electrical damping model fha: the time constant is adjustable to 0, 0.3, 1.2, 4.8, or 19.2 seconds. model fka: the time constant is adjustable between 0 to 38.4 seconds. (9 steps) zero elevation/suppression: zero can be elevated within the specified range limit of each sensor model. normal/reverse action: model fha: selectable by moving a jumper pin located on the electronics unit. model fka: configurable from hhc. indication: analog indicator or 4 -digit lcd meter, as specified. burnout direction: if self-diagnostic detect transmitter fail- ure, the analog signal will be driven to ei- ther ?utput hold? ?utput overscale or ?utput underscale?modes. model fha: unless otherwise specified in the order, the transmitter will be shipped in ?utput hold?mode. (output signal just before failure happens is maintained.) model fka: selectable from hhc ?utput hold? output signal is hold as the value just be- fore failure happens. performance specifications accuracy rating: (including linearity, hysteresis, and re- peatability). (standard) for spans greater than 1/10 of url: 0.2% of span for spans below 1/10 of url (model fka only): (0.1 + 0.1 )% of span 0.1 x url span note: (*) hhc's version must be more than 5.0 (or fxw 1 2), to use this function. 2000 [ ? ] 1500 1000 load resistance 500 250 1533 ? 600 ? operating area 24v 45v 10 16.1 50 [ v ] 10.5v power voltage baseefa ex ds iic t5, t6 eex ia iic t4, t5 ex n ii t5 factory class i ii iii class i ii iii class i ii iii mutual div. 1 div. 1 div. 2 groups b thru. g groups a thru. f groups a thru. g csa class i ii iii class i ii iii class i ii iii div. 1 div. 1 div. 2 groups c thru. g groups a thru. g groups a thru. g riis ex ds iib+h 2 t4 saa ex d ii c t5, t6 ex ia ii c t5, t6 ex n ii c t5, t6 ip 66/67 ip 66/67 ip 66/67 1 2
3 non-wetted parts material: electronics housing: low copper die-cast aluminum alloy (standard), finished with polyester coat ing, or 316 stainless steel (scs14 per jis g5121), as specified. bolts and nut: cr-mo alloy (standard) or 304 stainless steel fill fluid: silicone oil mounting bracket: carbon steel with epoxy coating or 304 stainless steel, as specified environmental protection: iec ip67 and nema 4x mounting: on 60.5mm (jis 50a) pipe using mounting bracket, direct wall mounting, or direct process mounting. mass{weight}: transmitter approximately 3.4kg without options. add; 0.5kg for mounting bracket 0.8kg for indicator option 4.5kg for stainless steel housing option notes: sensor o-rings: viton availability of above material design depends on ranges. refer to "code symbols". accessories optional features indicator: a plug-in analog indicator (1.5% accuracy) can be housed in the electronics compart- ment or in the terminal box of the housing. an optional 4 digits lcd meter is also available. arrester: a built-in arrester protects the electronics from lightning surges. lightning surge immunity: 4kv (1.2 50 s) degreasing: process-wetted parts are cleaned, but the fill fluid is standard silicone oil. not for use on oxygen or chlorine measurement. nace specification: metallic materials for all pressure bound- ary parts comply with nace mr-01-75. 304 stainless steel bolts and nuts, astm b7m or l7m bolts and 2hm nuts (class ii) are available. customer tag: a stainless steel tag for customer tag data is wired to the transmitter. coating of cell: cell's surface is finished with epoxy/ polyurethane double coating. specify if environment is extremely corrosive. (option) (not available for max span 16kpa abs, 130kpa abs) for spans greater than 1/10 of url: 0.1% of span for spans below 1/10 of url (model fka only): (0.05 + 0.05 )% of span linearity: 0.1% of calibrated span stability: 0.2% of upper range limit (url) for 24 months temperature effect: effect per 28 c change between the limits of 40 c and +85 c zero shift: (0.125+0.1 )% total effect: (0.15+0.1 )% double the effects for material code (7th digit in code symbols) "h", "m", and "t". overrange effect: zero shift; 0.2% of url for any overrange to maximum limit supply voltage effect: less than 0.05% of calibrated span per 10v rfi effect: less than 0.2% of url for the frequen- cies of 20 to 1000mhz and field strength 30v/m when electronics covers on. (classification: 2-abc: 0.2% span per sama pmc 33.1) step response: time constant. 0.2 s dead time: approximately 0.3 s (without electrical damping) mounting position effect: zero shift, less than 0.1kpa{1mbar} for a 10 tilt in any plane. no effect on span. this error can be cor- rected by adjusting zero. dielectric strength: 500v ac, 50/60hz 1 min., between circuit and earth. insulation resistance: more than 100m ? at 500v dc. turn-on time: 4 sec internal resistance for external field indicator: 12 ? or less physical specifications electrical connections: g1/2, 1/2-14npt, pg13.5, or m20 x 1.5 conduit, as specified. process connections: 1/4-18 npt or rc1/4, as specified. process-wetted parts material: oval flanges: (model ffp, refer to data sheet no. eds6-10) converts process connection to 1/2-14 npt or to rc1/2; in carbon steel or in 316 stainless steel. hand held communicator: (model fxw, refer to data sheet no.eds 8-47) communication module: (standard for model fka) by adding communication module, remote setting function becomes avail- able for fha. remark: when the communication module is connected, the operation mode of ex- ternal zero/span adjustment screw is limited to zero adjustment only. 1 2 url span url span 0.1 x url span v h m t material code (7th digit in "code symbols") process cover 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316l stainless steel hastelloy-c monel tantalum diaphragm 316 stainless steel hastelloy-c lining monel lining tantalum lining wetted sensor body 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel vent/drain
fha, fka ... 3 4 12345678 910111213 1415 description type 4 to 20ma, output type 4 to 20ma with digital signal, output type connections process connection rc1/4 1/4-18npt 1/4-18npt 1/4-18npt 1/4-18npt oval flange screw 7/16-20unf 7/16-20unf m10 m10 7/16-20unf conduit connection g 1/2 1/2-14npt pg 13.5 m20 1.5 pg 13.5 span limit span limit fha/fka [kpa abs]{bar abs}(*) 1.6/1.6?6/16 {0.016/0.016?.16/0.16 13/1.6?30/130 {0.13/0.016?.3/1.3} 50/5?00/500 {0.5/0.05?/5} 300/30?000/3000 {3/0.3?0/30} process cover 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316 stainless steel 316stainless steel 316stainless steel 316 stainless steel 316 stainless steel wetted cell body 316 stainless steel hast. c lining monel lining 316 stainless steel hast. c lining monel lining tantalum lining 316 stainless steel hast. c lining monel lining tantalum lining 316 stainless steel hast. c lining monel lining tantalum lining diaphragm 316l stainless steel hast. c monel 316l stainless steel hast. c monel tantalum 316l stainlesssteel hast. c monel tantalum 316l stainless steel hast. c monel tantalum 03 fha fka s t v w x note: * 100: 1 turn down is possible for model fka, but should be used at a span greater than 1/40 of the maximum span for better performance. 1v 1h 1m 2v 2h 2m 2t 3v 3h 3m 3t 4v 4h 4m 4t code symbols the product conforms to the requirements of the electromagnetic compatibility directive 89/336/eec as detailed within the technical construction file number tn510412. the applicable standards used to demonstrate compliance are :- emi (emission) en50081-1 : 1992 "lvd - the transmitter is not covered by the requirements of the lvd standard." ems (immunity) en50082-1 : 1992 test item frequency range basic standard applicable electro- magnetic radiation disturbance 30-1000mhz en55022 class b test item no. test specification basic standard performance criteria 1 2 3 b a b electrostatic discharge fast transients common mode 0.5kv, 5/50 (tr/th) ns 5khz rep. radio-frequency electromagnetic field. 8kv (air) 27-500mhz 3v/m (unmodulated) iec 801-2:1984 iec 801-3:1984 iec 801-4:1988
5 description 03 fka 03 fha 12345678 910111213 1415 indicator and arrester indicator arrester none none analog, 0 to 100% linear scale none analog, custom scale none analog, double scale none (jis and si unit) none yes analog, 0 to 100% linear scale yes analog, custom scale yes analog, double scale yes (jis and si unit) digital, 0 to 100% none digital, custom scale none (model fka only) digital, 0 to 100% yes digital, custom scale yes (model fka only) approvals for hazardous locations (approval pending) none (for ordinary locations) jis, flameproof (conduit seal) (available for 4th digit code ?? jis, flameproof (cable gland seal) (available for 4th digit code ?? fm, flameproof (or explosionproof) (available for 4th digit code ?? csa, flameproof (or explosionproof) (available for 4th digit code ?? baseefa, flameproof (conduit seal) baseefa, flameproof (cable gland seal) (conduit connection g 1/2 only) fm, intrinsic safety and nonincendive csa, intrinsic safety and nonincendive cenelec, intrinsic safety cenelec, intrinsic safety and baseefa, type n saa flameproof (conduit seal)(available for 4th digit code ?,t,w) saa intrinsic safety(available for 4th digit code ?,t,w) saa type?n(non-sparking)(available for 4th digit code ?,t,w) side vent/ drain and mounting bracket side vent/drain mounting bracket none none none yes, carbon steel none yes, stainless steel yes none yes yes, carbon steel yes yes, stainless steel stainless steel parts stainless steel tag plate stainless steel elec. housing coating of cell none none none yes none none none yes none yes yes none none none yes yes none yes none yes yes yes yes yes special applications and fill fluid treatment fill fluid none (standard) silicone oil degreasing silicone oil nace specification silicone oil (7th digit code "t" and 15th digit code "a", "b" are not available) sensor o-ring viton bolt/nut cr-mo alloy hexagon socket head cap screw/carbon steel nut cr-mo alloy hexagon bolt/nut nace bolt/nut (astm a193 b7m/a194 2hm) nace bolt/nut (astm a320 l7m/a194 2hm) 304 stainless steel/304 stainless steel (*1) * (1) in case of tropical use, select stainless bolts and nuts. a b c d e m n h j k p r t q a b c d e f y b c e m n p q y g n a b d j e f h k l p q s a b c d e a
fha, fka ... 3 information in this catalog is subject to change without notice. printed in japan head office 11-2 osaki 1-chome, shinagawa-ku, tokyo, 141-0032 japan phone: 81-3-5435-7111 http://www.fujielectric.co.jp/eng/sg/keisoku/welcome.htm sales div. international sales dept. no.1, fuji-machi, hino-city, tokyo, 191-8502 japan phone: 81-42-585-6201, 6202 fax: 81-42-585-6187, 6189 ?8.5 h 58 30 224 27 70 120 252 u - bolt m8 vent/drain plug 4 holes 9 dia. 17 g 1/2 ?3 d e f ?5 ? see table 1 see table 1 side vent (option) up down 41.3 50 2.5 8 2-m8 12 detail ? detail ? (conduit conn.) (press conn.) detail ? table 1 4th digit code conduit conn. d ef press. conn. 17 8 earth terminal m4 m4 1/2 ?4npt 16 5 oval flange screw 7/16-20unf screw depth13 h g1/2 pg13.5 m20 1.5 pg13.5 8 16 8 4.5 5 4.5 rc1/4 1/4?8npt 1/4?8npt 1/4?8npt 1/4?8npt no. 8-32unc 7/16-20unf screw depth13 m10 screw depth13 7m10 screw depth13 7/16-20unf screw depth13 m4 m4 t x s v w cable gland (option) see note ? conduit conn. ? conduit conn. fastener (in case of 10th digit code ?,c,m,n? 104 ?0 43.5 126.5 zero/span adj. screw 51 71 ?0 (27) with field ind. (option) oval flange screw see table1 1 mtg. pipe jis 50a (2") ? press. conn. vent/drain plug (when the vent/drain plug attatched on the upper side of the detector) 110 50 88 in case of 4th digit code ?,v,w,x + connection diagram + ck+ ck 38 70 note1) : cable gland is supplied in case of flamproof packing type. ?1 cable is suitable. earth terminal (both side ) see table1 f h k a s t v w x 0 v h m t 3 outline diagram (unit:mm)


▲Up To Search▲   

 
Price & Availability of EDSX5-71

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X